www.rkiinstruments.com61-1007RKSSCO2 DetectorOperator’s ManualPart Number: 71-0297RKRevision: 0Released: 7/18/14
6 • 61-1007RKSS Carbon Dioxide Gas DetectorFigure 4: Wiring the CO2 Detector to a Controller9. If using shielded cable, connect the cable’s drain wi
61-1007RKSS Carbon Dioxide Gas Detector • 7Start UpThis section describes procedures to start up the CO2 detector and place the detector into normal
8 • 61-1007RKSS Carbon Dioxide Gas DetectorMaintenanceThis section describes maintenance procedures. It includes preventive maintenance, troubleshoot
61-1007RKSS Carbon Dioxide Gas Detector • 9NOTE: If the reading is not within ± 20% of the gas concentration, calibrate the detector as described in
10 • 61-1007RKSS Carbon Dioxide Gas DetectorReplacing Components of the CO2 DetectorThis section includes a procedure to replace the hydrophobic memb
61-1007RKSS Carbon Dioxide Gas Detector • 11lose the cap gasket.4. Unplug and remove the plug-in IR CO2 sensor.5. Carefully plug the replacement sens
12 • 61-1007RKSS Carbon Dioxide Gas DetectorCalibration FrequencyAlthough there is no particular calibration frequency that is correct for all applic
61-1007RKSS Carbon Dioxide Gas Detector • 13allow the gas to flow to the detector for 1 minute.3. Turn the regulator’s on/off knob clockwise to close
14 • 61-1007RKSS Carbon Dioxide Gas DetectorParts ListTable 3 lists replacement parts and accessories for the 61-1007RKSS CO2 detector.Table 3: Parts
61-1007RKSS Carbon Dioxide Gas DetectorWARNINGRead and understand this instruction manual before operating detector. Improper use of the detector cou
61-1007RKSS Carbon Dioxide Gas DetectorProduct WarrantyRKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in mater
61-1007RKSS Carbon Dioxide Gas DetectorTable of ContentsOverview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
61-1007RKSS Carbon Dioxide Gas Detector • 1OverviewThis manual describes the 61-1007RKSS CO2 (carbon dioxide) detector. This manual also describes ho
2 • 61-1007RKSS Carbon Dioxide Gas DetectorDescriptionThe 61-1007RKSS CO2 detector has four versions, the 61-1007RKSS-02 which has a detection range
61-1007RKSS Carbon Dioxide Gas Detector • 3Infrared CO2 DetectorThe CO2 detector consists of the detector housing body, detector housing cap, cap gas
4 • 61-1007RKSS Carbon Dioxide Gas Detectorconnect the wiring from the detector to a controller. An O-ring seals the interface between the junction b
61-1007RKSS Carbon Dioxide Gas Detector • 52. At the mounting site you select, hang or mount the junction box with the detector facing down (see Figu