RKI Instruments, Inc. • 1855 Whipple Rd. • Hayward, CA 94544 • (510) 441-5656 Model GD-K11D Operator’s Manual Part Number: 71-0029RKEdition: FirstR
6 • Description Model GD-K11D Operator’s Manual Status lights The display board includes five status lights. Table 3 lists each light and its functi
Model GD-K11D Operator’s Manual Description • 7 Adjustment potentiometers The display board includes two potentiometers, which extend from the bott
8 • Installation Model GD-K11D Operator’s Manual Installation This section describes how to install the Model GD-K11D at the monitoring site. This
Model GD-K11D Operator’s Manual Installation • 9 2. Use two screws through the slot cutouts at the top of the housing to secure the housing to a ve
10 • Installation Model GD-K11D Operator’s Manual Connecting AC power 1. Loosen the thumbscrew at the top of the housing.2. Rotate the housing cove
Model GD-K11D Operator’s Manual Start Up • 11 Start Up This section describes procedures to start up the Model GD-K11D and place the Model GD-K11D
12 • Start Up Model GD-K11D Operator’s Manual6. Press the MODE button.The ZERO light turns off, and 3. appears on the display screen.NOTE: The purpo
Model GD-K11D Operator’s Manual Operation • 13OperationThis section describes procedures to display the peak reading, and to adjust the alarm point,
14 • Operation Model GD-K11D Operator’s Manual3. Adjust the INT switch to the “2” position.The current alarm point appears on the display screen, and
Model GD-K11D Operator’s Manual Operation • 15Adjusting the Alarm TypeYou can program the Model GD-K11D’s alarm to activate on increasing readings or
ii • Warranty Model GD-K11D Operator’s Manual Warranty RKI Instruments, Inc., warranties gas alarm equipment manufactured by RKI and sold by RKI to
16 • Alarms Model GD-K11D Operator’s ManualAlarmsThis section describes the Model GD-K11D’s visual indications for normal, alarm, and fail conditions
Model GD-K11D Operator’s Manual Maintenance • 17MaintenanceThis section describes maintenance procedures for the Model GD-K11D. It includes preventiv
18 • Maintenance Model GD-K11D Operator’s ManualTroubleshootingThe troubleshooting table describes error messages, symptoms, probable causes, and rec
Model GD-K11D Operator’s Manual Maintenance • 194. Pull the metal restraining bracket that secures the sensor in the housing away from the sensor, th
20 • Maintenance Model GD-K11D Operator’s Manual4. Use the ZERO potentiometer to adjust the display reading to zero.5. Press the MODE button.The ZERO
Model GD-K11D Operator’s Manual Calibration • 21CalibrationThis section describes how to calibrate the Model GD-K11D. It includes procedures to prepa
22 • Calibration Model GD-K11D Operator’s ManualSetting the Response (Span) Reading1. Adjust the INT switch to the “1” position. The INT switch is th
Model GD-K11D Operator’s Manual Table of Contents • iii Table of Contents Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
iv Model GD-K11D Operator’s Manual
Model GD-K11D Operator’s Manual Overview • 1 Overview The Model GD-K11D sample-draw transmitter is a single-point, fixed-instrument. It continuously
2 • Specifications Model GD-K11D Operator’s Manual Table 2 lists the gases that the Model GD-K11D is capable of detecting. Table 2 also lists the d
Model GD-K11D Operator’s Manual Description • 3 Description This section describes the components of the Model GD-K11D. The Model GD-K11D consists
4 • Description Model GD-K11D Operator’s Manual Electrical System The Model GD-K11D’s electrical system consists of the terminal strip, power switc
Model GD-K11D Operator’s Manual Description • 5 Detection System The Model GD-K11D’s detection system consists of the sensor, lithium battery, ampl