Rki-instruments GX-8000(PT0-098) Manual de usuario Pagina 39

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 52
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 38
6 Maintenance 6-1. Maintenance intervals and items
- 39 - GX-8000
6
Maintenance
This is an important instrument for the purpose of safety.
To maintain the performance of the gas monitor and improve the reliability of safety, perform a regular
maintenance.
6-1. Maintenance intervals and items
Daily maintenance: Perform maintenance before beginning to work.
Monthly maintenance: Perform alarm test once a month.
Regular maintenance: Perform a maintenance once or more for every six months to maintain the
performance as a safety unit.
Maintenance item Maintenance content
Daily
maintenance
Monthly
maintenance
Regular
maintenance
Battery Level
Check
Check that the battery level is sufficient.
{ { {
Concentration
Display Check
Make the gas monitor draw in fresh air and check that
the concentration display value is zero (or 20.9 vol%
on the oxygen deficiency meter). When the reading is
incorrect, perform the zero adjustment (fresh air
adjustment) after ensuring that no other gases exist
around it.
{ { {
Flow Rate Check
See the flow rate indicator to check for abnormalities.
{ { {
Filter Check
Check the dust filter for dust or clogging.
{ { {
Alarm Test Check the alarm lamp and buzzer for normal
operation using the alarm test function.
{ {
Span Adjustment Perform the span adjustment by using the calibration
gas.
{
Gas Alarm Check
Check the gas alarm by using the calibration gas.
{
Vista de pagina 38
1 2 ... 34 35 36 37 38 39 40 41 42 43 44 ... 51 52

Comentarios a estos manuales

Sin comentarios