81 • Calibrating a Gas Tracer
If the ppm sensor fails the fresh air adjustment, the SDM-2012 will abort
the calibration and will not apply calibration gas. If this happens, the
CAL S LED will flash red indicating a failure and the following screens will
alternate.
In this case continue with step 28.
19. The SDM-2012 will apply GAS 2 calibration gas to the ppm sensor for the time
defined by the GAS TIME calibration parameter. Be sure that the 1000 ppm
methane calibration cylinder is connected to the GAS 2 fitting.
20. The SDM-2012 will then apply fresh air to the standard three sensors and
%volume sensor (if installed and if the GAS menu item in Edit Mode is set to
ALL) for the time defined by the AIR TIME calibration parameter.
21. The SDM-2012 will perform a fresh air adjustment on the standard three
sensors and %volume sensor (if installed and if the GAS menu item in Edit
Mode is set to ALL).
If the air adjustment is successful, the SDM-2012 will continue as
described in step 22 below.