50 • Calibratin
g
the Model GX-2001 with the Calibration Station
9. To begin calibration, click the “Cal. Unit Calibration” button on
the Download window.
10.After the Model GX-2001 has been calibrated, remove it
from the Charger/Downloading Station by pressing down on
the metal release lever located on the right side of the
Calibration Adapter.
11.Place another Model GX-2001 in the Data Logging/Charger
Station if you have other unit(s) to calibrate. Repeat steps 5 –
10 above.
12.When you are done calibrating, remove the regulator from
the gas cylinder.
13.Disconnect the regulator from the Span Gas In tubing. Store
the regulator and gas cylinder in a safe place.
14.Turn off the Calibration Station.
Figure 30. Choosing the Cal. Unit Calibration Option