Rki-instruments SDM-E2 Manual de usuario Pagina 66

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 89
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 65
61 • Calibrating an Instrument with Special Sensors
release the EDIT ENTER button on the SDM-E2. The calibration will continue.
19. Once the GAS 2 calibration gas has been applied for the appropriate amount
of time, the SDM-E2 will determine if the sixth channel passed the calibration.
20. The SDM-E2 will then purge the system with fresh air for the time defined by
the AIR TIME calibration parameter. The calibration results will be displayed
with a P (pass) or an F (fail) while a fresh air purge is performed.
21. Once the fresh air purge is finished, the CAL S LED will stop blinking and be
steadily green if the calibration passed or steadily red if the calibration failed.
22. After a successful or failed calibration,
To perform any other operations:
To perform a bump test, press and hold the BUMP T button until the pump
starts. To perform another calibration, press and hold the CAL S button
until the pump starts. To return to the home screen, press and hold the
CH4 29%LEL CF
OXY 18.0vol% CF
H2S 25.0ppm CP
CO 45ppm CP
CO2 3500ppm CF
NH3 3.0ppm C
CH4 29%LEL CF
OXY 18.0vol% CF
H2S 25.0ppm CP
CO 45ppm CP
CO2 3500ppm CF
NH3 3.0ppm CF
CH4 29%LEL CF
OXY 18.0vol% CF
H2S 25.0ppm CP
CO 45ppm CP
CO2 3500ppm CF
NH3 3.0ppm CF
Vista de pagina 65
1 2 ... 61 62 63 64 65 66 67 68 69 70 71 ... 88 89

Comentarios a estos manuales

Sin comentarios