Rki-instruments SDM-E2 Manual de usuario

Busca en linea o descarga Manual de usuario para Sensores Rki-instruments SDM-E2. RKI Instruments SDM-E2 User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir

Indice de contenidos

Pagina 1 - Operator’s Manual

www.rkiinstruments.com SDM-E2 Docking StationStandalone ConfigurationOperator’s ManualPart Number: 71-0224RKRevision: DReleased: 11/19/13

Pagina 2 - Warranty

5 • OverviewChapter 2: DescriptionOverviewThis section describes the SDM-E2 docking station. It is designed to be used on a table top and consists of

Pagina 3 - Table of Contents

USB Cable • 6USB CableA Type A to Type B USB cable is provided with the docking station. It is only for use with the PC Controlled configuration. It

Pagina 4

7 • Instrument PanelA check valve is included with the SDM-E2 but is not needed for the Standalone configuration. It is used for the PC Controlled co

Pagina 5

Back Panel • 8Back PanelThe back panel includes the power jack, sample fittings, gas out to EAGLE 2 fitting, and a USB PC connector. Power JackThe po

Pagina 6 - Chapter 1: Introduction

9 • Control PanelControl PanelThe control panel is used to setup and operate the docking station in the Standalone configuration. It is located at th

Pagina 7 - System Requirements

Control Panel • 10CAL S • Initiates a calibration• Cancels a calibration• Clears data from docking station memory (when used with COPY button)• Move

Pagina 8 - Specifications

11 • Front PanelFront PanelA type A USB port is located on the front of the docking station. This port can be used to save calibration and bump test

Pagina 9 - About this Manual

Overview • 12Chapter 3: Preparing to Use the SDM-E2OverviewThere are four tasks that must be completed before you can begin to use the SDM-E2: hardwa

Pagina 10 - Chapter 2: Description

13 • Setting the Operational Parameters in Edit Modesafely dispersed, such as an open window. CAUTION: The maximum recommended length for the exhaust

Pagina 11 - USB Cable

Setting the Operational Parameters in Edit Mode • 14Air Sample Time (AIR TIME)The air sample time can be set separately for bump testing and calibrat

Pagina 12 - Instrument Panel

WarrantyWarrantyRKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials and workmanship, and performance f

Pagina 13 - Back Panel

15 • Setting the Operational Parameters in Edit ModeTurning on the SDM-E2 with an EAGLE 2Do the following to turn on the SDM-E2 and establish a conne

Pagina 14 - Control Panel

Setting the Operational Parameters in Edit Mode • 16 6. If the EAGLE 2 is equipped with NiMH batteries and you wish to charge the batteries, connect

Pagina 15

17 • Setting the Operational Parameters in Edit ModeSetting the Bump Test ParametersDo the following to set the bump test parameters after establishi

Pagina 16 - Front Panel

Setting the Operational Parameters in Edit Mode • 18 4. If you wish to cancel setting the bump test parameters or were just viewing the parameters to

Pagina 17 - Hardware Assembly

19 • Setting the Operational Parameters in Edit ModeSetting the Calibration ParametersDo the following to set the calibration parameters after establ

Pagina 18

Setting the Operational Parameters in Edit Mode • 20 4. If you wish to cancel setting the calibration parameters or were just viewing the parameters

Pagina 19

21 • Setting the Operational Parameters in Edit ModeSetting the Gas Inlet ParametersThe gas inlet parameters define what gas ports are used during bu

Pagina 20 - C a li bra ti on S ta tio n

Setting the Operational Parameters in Edit Mode • 22 3. Press and release the EDIT ENTER button. The EAGLE 2 will display the following screen with t

Pagina 21

23 • Setting the Operational Parameters in Edit Mode 7. To edit the gas assignments, move the cursor next to the ASSIGN parameter with the BUMP T or

Pagina 22

Connecting Calibration Gas • 24Connecting Calibration Gas The GAS 1 and GAS 2 fittings on the back of the docking station are designed to be used wit

Pagina 23

Table of ContentsTable of ContentsChapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 24

25 • Connecting Calibration GasUse Table 4 below as a guide in determining which calibration gas cylinder is appropriate for your system. See “Spare

Pagina 25

Installing the Single Module Data Viewer Software • 26Installing the Single Module Data Viewer Software 1. Launch Windows®. 2. Exit from all applicat

Pagina 26

27 • OverviewChapter 4: OperationOverviewWhen you have completed the tasks in "Chapter 3: Preparing to Use the SDM-E2", you are ready to us

Pagina 27

Bump Testing Instruments with Standard Sensors • 28 5. Install an EAGLE 2 in the instrument cradle. Set the EAGLE 2 onto the top of the SDM-E2 and sl

Pagina 28

29 • Bump Testing Instruments with Standard SensorsWARNING: Do not plug the charger cable into a battery pack that contains alkaline batteries. Do no

Pagina 29 - Connecting Calibration Gas

Bump Testing Instruments with Standard Sensors • 30 11. Press and hold the BUMP T button until the BUMP T LED turns on (about one second) then releas

Pagina 30

31 • Bump Testing Instruments with Standard Sensors 14. The SDM-E2 will perform a fresh air adjustment on the instrument.• If the air adjustment is s

Pagina 31 - Software

Bump Testing Instruments with Standard Sensors • 32 15. The SDM-E2 will apply calibration gas to the instrument for the time defined by the GAS TIME

Pagina 32 - Chapter 4: Operation

33 • Bump Testing Instruments with Standard Sensors 17. When the AUTO CAL bump test parameter is set to ON:• If all channels pass the bump test: a. T

Pagina 33 - C a libra ti on St atio n

Bump Testing Instruments with Standard Sensors • 34The current gas readings will be displayed along with the bump test results. A “C” will appear nex

Pagina 34

Table of ContentsSetting the Gas Inlet Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Connecting

Pagina 35

35 • Calibrating Instruments with Standard Sensorsseconds. If the bump test failed, the instrument will shut off after 10 minutes. If buttons are pre

Pagina 36

Calibrating Instruments with Standard Sensors • 36To perform a calibration on an instrument: 1. Confirm that the AC Adapter is connected to the SDM-E

Pagina 37

37 • Calibrating Instruments with Standard Sensors 6. If the EAGLE 2 is equipped with NiMH batteries and you wish to charge the batteries, connect th

Pagina 38

Calibrating Instruments with Standard Sensors • 38 10. Press and hold the CAL S button until the CAL S LED turns on (about one second) then release i

Pagina 39

39 • Calibrating Instruments with Standard Sensors 13. The SDM-E2 will perform a fresh air adjustment on the instrument.• If the air adjustment is su

Pagina 40

Calibrating Instruments with Standard Sensors • 40with a P (pass) or an F (fail) while the fresh air purge is performed. 16. Once the fresh air purge

Pagina 41 - C ali bra ti on S t at io n

41 • Bump Testing an Instrument with Special Sensorssaved bump test and calibration records to a USB flash drive. 19. Disconnect the charger cable,

Pagina 42

Bump Testing an Instrument with Special Sensors • 42 5. Install an EAGLE 2 in the instrument cradle. Set the EAGLE 2 onto the top of the SDM-E2 and s

Pagina 43

43 • Bump Testing an Instrument with Special SensorsWARNING: Do not plug the charger cable into a battery pack that contains alkaline batteries. Do n

Pagina 44

Bump Testing an Instrument with Special Sensors • 44NOTE: If you have an EAGLE 2 with an IR CO2 sensor in the 0-10,000 ppm or 0-5.00 %vol range, be s

Pagina 45

Table of ContentsChanging the Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79Exiting

Pagina 46 - Bump Testing an Instrument

45 • Bump Testing an Instrument with Special Sensors• To perform a calibration, press and release the CAL S button. 13. The “Z” to the right of the g

Pagina 47

Bump Testing an Instrument with Special Sensors • 46 15. The SDM-E2 will apply GAS 1 calibration gas to the instrument for the time defined by the GA

Pagina 48

47 • Bump Testing an Instrument with Special Sensorsb. Once the GAS 2 calibration gas has been applied for the appropriate amount of time, the SDM-E2

Pagina 49

Bump Testing an Instrument with Special Sensors • 48e. Once the GAS 2 calibration gas has been applied for the appropriate amount of time, the SDM-E2

Pagina 50

49 • Bump Testing an Instrument with Special Sensorsb. GAS 2 calibration gas will be applied to the fifth sensor and the SDM-E2 will determine if it

Pagina 51

Bump Testing an Instrument with Special Sensors • 50e. Once the GAS 2 calibration gas has been applied for the appropriate amount of time, the SDM-E2

Pagina 52

51 • Bump Testing an Instrument with Special SensorsNOTE: GAS 1 calibration gas will continue to be applied so that the total gas application time is

Pagina 53

Bump Testing an Instrument with Special Sensors • 52d. Once the calibration is complete, the screen will indicate whether the fifth channel calibrati

Pagina 54

53 • Bump Testing an Instrument with Special Sensorsg. GAS 2 calibration gas will be applied to the sixth channel and the SDM-E2 will determine if it

Pagina 55

Bump Testing an Instrument with Special Sensors • 54instrument, it will automatically turn it off 10 minutes after the last button push.To turn off t

Pagina 56

1 • OverviewChapter 1: IntroductionOverviewThis chapter briefly describes the SDM-E2 Docking Station and the Single Module Data Viewer Program. This

Pagina 57

55 • Calibrating an Instrument with Special SensorsCalibrating an Instrument with Special SensorsWhen a calibration is performed, the docking station

Pagina 58

Calibrating an Instrument with Special Sensors • 56 5. Install an EAGLE 2 in the instrument cradle. Set the EAGLE 2 onto the top of the SDM-E2 and sl

Pagina 59

57 • Calibrating an Instrument with Special SensorsWARNING: Do not plug the charger cable into a battery pack that contains alkaline batteries. Do no

Pagina 60 - Calibrating an Instrument

Calibrating an Instrument with Special Sensors • 58 10. Press and hold the CAL S button until the CAL S LED turns on (about one second) then release

Pagina 61 - Calibration Station

59 • Calibrating an Instrument with Special Sensors 13. The SDM-E2 will perform a fresh air adjustment on the instrument.• If the air adjustment is s

Pagina 62

Calibrating an Instrument with Special Sensors • 60calibrated yet. 15. The calibration results for the standard 4 channels will be displayed by a P (

Pagina 63

61 • Calibrating an Instrument with Special Sensorsrelease the EDIT ENTER button on the SDM-E2. The calibration will continue. 19. Once the GAS 2 cal

Pagina 64

Calibrating an Instrument with Special Sensors • 62EDIT ENTER button until the home screen appears. • To turn the EAGLE 2 off:If the calibration was

Pagina 65

63 • TroubleshootingTroubleshootingNOTE: This troubleshooting guide describes SDM-E2 problems only. See the EAGLE 2 Operator’s Manual for problems yo

Pagina 66

Troubleshooting • 64• No IR connection between EAGLE 2 and SDM-E2• SDM-E2 is not turned on.• The EAGLE 2 is not correctly inserted into the instrumen

Pagina 67

System Requirements • 2CAUTION: The EAGLE 2 detects oxygen deficiency and elevated levels of oxygen, combustible gases, carbon monoxide, and hydrogen

Pagina 68 - Troubleshooting

65 • Charging an Instrument in a Docking StationCharging an Instrument in a Docking StationThe SDM-E2 can be used to charge the rechargeable NiMH bat

Pagina 69 - Troubleshooting • 64

Charging an Instrument in a Docking Station • 66NOTE: If alkaline batteries are detected, the SDM-E2 must be turned off and on again before attemptin

Pagina 70 - Bump Test or Calibration

67 • Charging an Instrument in a Docking StationBatteries Too Drained for Standalone OperationIf the charger is connected to the EAGLE 2 and the batt

Pagina 71

Calibration and Bump Test Records • 68Calibration and Bump Test RecordsThe SDM-E2 saves a record of each bump test and calibration performed. It is c

Pagina 72 - WHILE CHARGING

69 • Calibration and Bump Test RecordsSDM-E2 will take a few seconds to determine how much memory is available in the flash drive. • If the flash dri

Pagina 73

Calibration and Bump Test Records • 70The files that an SDM-E2 saves to a USB flash drive are structured so that they can be imported into a database

Pagina 74 - Clearing the SDM-E2’s Memory

71 • OverviewChapter 5: Single Module Data Viewer ProgramOverviewThe Single Module Data Viewer Program is used to view, organize, and print bump test

Pagina 75 - Multiple Times In One Day

Using the Single Module Data Viewer Program • 72NOTE: When you start the Single Module Data Viewer Program for the first time, there will be no data

Pagina 76

73 • Using the Single Module Data Viewer Program• View the bump test and calibration data saved in the database.• Delete data.• Print bump test or ca

Pagina 77

Using the Single Module Data Viewer Program • 74Organizing the Data When viewing the data, it can be organized in two ways: 1. Base View FormatIn bas

Pagina 78

3 • SpecificationsSpecificationsTable 1: SDM-E2 SpecificationsInput Power 12 VDCNOTE: AC Adapter with 100 - 240 VAC, 50/60 Hz, 0.6A input and 12 VDC,

Pagina 79 - Organizing the Data

75 • Using the Single Module Data Viewer Program 2. ID View FormatIn ID view format, the data can be organized by one or more of the following items

Pagina 80 - Viewing the Data

Using the Single Module Data Viewer Program • 76 2. When an item no longer has a (+) or (-) symbol next to it, single click it and the contents of th

Pagina 81

77 • Using the Single Module Data Viewer ProgramIf you organize the data by date, then folders organized by year/month appear in the left side of the

Pagina 82

Using the Single Module Data Viewer Program • 78If you are viewing data in ID view format, expand the folders in the left side of the window until th

Pagina 83

79 • Using the Single Module Data Viewer Programappear showing the selected printer and confirming if you want to print. Verify that the selected pri

Pagina 84 - Changing the Password

Spare Parts List • 80window will appear prompting you for the new password again. 5. Enter the new password again and click the Confirm New Password

Pagina 85 - Spare Parts List

81 • Spare Parts List81-0018RK-01 Calibration cylinder, 50 %LEL isobutane in air, 34 liter steel81-0018RK-03 Calibration cylinder, 50 %LEL isobutane

Pagina 86

Spare Parts List • 8281-0142RK-04 5-gas calibration cylinder, CH4/O2/H2S/CO/SO2, 34 liter aluminum81-0143RK-02 Five-gas calibration cylinder, CH4/O2/

Pagina 87

83 • Appendix A: Bump Testing and Calibrating EAGLE 2s with IR SensorsAppendix A: Bump Testing and Calibrating EAGLE 2s with IR SensorsThe default ga

Pagina 88 - EAGLE 2s with IR Sensors

Appendix A: Bump Testing and Calibrating EAGLE 2s with IR Sensors • 84An IR sensor that can be bump tested or calibrated with one or more of the stan

Pagina 89

About this Manual • 4About this ManualThe SDM-E2 Docking Station Standalone Configuration Operator’s Manual uses the following conventions for notes,

Comentarios a estos manuales

Sin comentarios