www.rkiinstruments.com SDM-E2 Docking StationStandalone ConfigurationOperator’s ManualPart Number: 71-0224RKRevision: DReleased: 11/19/13
5 • OverviewChapter 2: DescriptionOverviewThis section describes the SDM-E2 docking station. It is designed to be used on a table top and consists of
USB Cable • 6USB CableA Type A to Type B USB cable is provided with the docking station. It is only for use with the PC Controlled configuration. It
7 • Instrument PanelA check valve is included with the SDM-E2 but is not needed for the Standalone configuration. It is used for the PC Controlled co
Back Panel • 8Back PanelThe back panel includes the power jack, sample fittings, gas out to EAGLE 2 fitting, and a USB PC connector. Power JackThe po
9 • Control PanelControl PanelThe control panel is used to setup and operate the docking station in the Standalone configuration. It is located at th
Control Panel • 10CAL S • Initiates a calibration• Cancels a calibration• Clears data from docking station memory (when used with COPY button)• Move
11 • Front PanelFront PanelA type A USB port is located on the front of the docking station. This port can be used to save calibration and bump test
Overview • 12Chapter 3: Preparing to Use the SDM-E2OverviewThere are four tasks that must be completed before you can begin to use the SDM-E2: hardwa
13 • Setting the Operational Parameters in Edit Modesafely dispersed, such as an open window. CAUTION: The maximum recommended length for the exhaust
Setting the Operational Parameters in Edit Mode • 14Air Sample Time (AIR TIME)The air sample time can be set separately for bump testing and calibrat
WarrantyWarrantyRKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials and workmanship, and performance f
15 • Setting the Operational Parameters in Edit ModeTurning on the SDM-E2 with an EAGLE 2Do the following to turn on the SDM-E2 and establish a conne
Setting the Operational Parameters in Edit Mode • 16 6. If the EAGLE 2 is equipped with NiMH batteries and you wish to charge the batteries, connect
17 • Setting the Operational Parameters in Edit ModeSetting the Bump Test ParametersDo the following to set the bump test parameters after establishi
Setting the Operational Parameters in Edit Mode • 18 4. If you wish to cancel setting the bump test parameters or were just viewing the parameters to
19 • Setting the Operational Parameters in Edit ModeSetting the Calibration ParametersDo the following to set the calibration parameters after establ
Setting the Operational Parameters in Edit Mode • 20 4. If you wish to cancel setting the calibration parameters or were just viewing the parameters
21 • Setting the Operational Parameters in Edit ModeSetting the Gas Inlet ParametersThe gas inlet parameters define what gas ports are used during bu
Setting the Operational Parameters in Edit Mode • 22 3. Press and release the EDIT ENTER button. The EAGLE 2 will display the following screen with t
23 • Setting the Operational Parameters in Edit Mode 7. To edit the gas assignments, move the cursor next to the ASSIGN parameter with the BUMP T or
Connecting Calibration Gas • 24Connecting Calibration Gas The GAS 1 and GAS 2 fittings on the back of the docking station are designed to be used wit
Table of ContentsTable of ContentsChapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
25 • Connecting Calibration GasUse Table 4 below as a guide in determining which calibration gas cylinder is appropriate for your system. See “Spare
Installing the Single Module Data Viewer Software • 26Installing the Single Module Data Viewer Software 1. Launch Windows®. 2. Exit from all applicat
27 • OverviewChapter 4: OperationOverviewWhen you have completed the tasks in "Chapter 3: Preparing to Use the SDM-E2", you are ready to us
Bump Testing Instruments with Standard Sensors • 28 5. Install an EAGLE 2 in the instrument cradle. Set the EAGLE 2 onto the top of the SDM-E2 and sl
29 • Bump Testing Instruments with Standard SensorsWARNING: Do not plug the charger cable into a battery pack that contains alkaline batteries. Do no
Bump Testing Instruments with Standard Sensors • 30 11. Press and hold the BUMP T button until the BUMP T LED turns on (about one second) then releas
31 • Bump Testing Instruments with Standard Sensors 14. The SDM-E2 will perform a fresh air adjustment on the instrument.• If the air adjustment is s
Bump Testing Instruments with Standard Sensors • 32 15. The SDM-E2 will apply calibration gas to the instrument for the time defined by the GAS TIME
33 • Bump Testing Instruments with Standard Sensors 17. When the AUTO CAL bump test parameter is set to ON:• If all channels pass the bump test: a. T
Bump Testing Instruments with Standard Sensors • 34The current gas readings will be displayed along with the bump test results. A “C” will appear nex
Table of ContentsSetting the Gas Inlet Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Connecting
35 • Calibrating Instruments with Standard Sensorsseconds. If the bump test failed, the instrument will shut off after 10 minutes. If buttons are pre
Calibrating Instruments with Standard Sensors • 36To perform a calibration on an instrument: 1. Confirm that the AC Adapter is connected to the SDM-E
37 • Calibrating Instruments with Standard Sensors 6. If the EAGLE 2 is equipped with NiMH batteries and you wish to charge the batteries, connect th
Calibrating Instruments with Standard Sensors • 38 10. Press and hold the CAL S button until the CAL S LED turns on (about one second) then release i
39 • Calibrating Instruments with Standard Sensors 13. The SDM-E2 will perform a fresh air adjustment on the instrument.• If the air adjustment is su
Calibrating Instruments with Standard Sensors • 40with a P (pass) or an F (fail) while the fresh air purge is performed. 16. Once the fresh air purge
41 • Bump Testing an Instrument with Special Sensorssaved bump test and calibration records to a USB flash drive. 19. Disconnect the charger cable,
Bump Testing an Instrument with Special Sensors • 42 5. Install an EAGLE 2 in the instrument cradle. Set the EAGLE 2 onto the top of the SDM-E2 and s
43 • Bump Testing an Instrument with Special SensorsWARNING: Do not plug the charger cable into a battery pack that contains alkaline batteries. Do n
Bump Testing an Instrument with Special Sensors • 44NOTE: If you have an EAGLE 2 with an IR CO2 sensor in the 0-10,000 ppm or 0-5.00 %vol range, be s
Table of ContentsChanging the Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79Exiting
45 • Bump Testing an Instrument with Special Sensors• To perform a calibration, press and release the CAL S button. 13. The “Z” to the right of the g
Bump Testing an Instrument with Special Sensors • 46 15. The SDM-E2 will apply GAS 1 calibration gas to the instrument for the time defined by the GA
47 • Bump Testing an Instrument with Special Sensorsb. Once the GAS 2 calibration gas has been applied for the appropriate amount of time, the SDM-E2
Bump Testing an Instrument with Special Sensors • 48e. Once the GAS 2 calibration gas has been applied for the appropriate amount of time, the SDM-E2
49 • Bump Testing an Instrument with Special Sensorsb. GAS 2 calibration gas will be applied to the fifth sensor and the SDM-E2 will determine if it
Bump Testing an Instrument with Special Sensors • 50e. Once the GAS 2 calibration gas has been applied for the appropriate amount of time, the SDM-E2
51 • Bump Testing an Instrument with Special SensorsNOTE: GAS 1 calibration gas will continue to be applied so that the total gas application time is
Bump Testing an Instrument with Special Sensors • 52d. Once the calibration is complete, the screen will indicate whether the fifth channel calibrati
53 • Bump Testing an Instrument with Special Sensorsg. GAS 2 calibration gas will be applied to the sixth channel and the SDM-E2 will determine if it
Bump Testing an Instrument with Special Sensors • 54instrument, it will automatically turn it off 10 minutes after the last button push.To turn off t
1 • OverviewChapter 1: IntroductionOverviewThis chapter briefly describes the SDM-E2 Docking Station and the Single Module Data Viewer Program. This
55 • Calibrating an Instrument with Special SensorsCalibrating an Instrument with Special SensorsWhen a calibration is performed, the docking station
Calibrating an Instrument with Special Sensors • 56 5. Install an EAGLE 2 in the instrument cradle. Set the EAGLE 2 onto the top of the SDM-E2 and sl
57 • Calibrating an Instrument with Special SensorsWARNING: Do not plug the charger cable into a battery pack that contains alkaline batteries. Do no
Calibrating an Instrument with Special Sensors • 58 10. Press and hold the CAL S button until the CAL S LED turns on (about one second) then release
59 • Calibrating an Instrument with Special Sensors 13. The SDM-E2 will perform a fresh air adjustment on the instrument.• If the air adjustment is s
Calibrating an Instrument with Special Sensors • 60calibrated yet. 15. The calibration results for the standard 4 channels will be displayed by a P (
61 • Calibrating an Instrument with Special Sensorsrelease the EDIT ENTER button on the SDM-E2. The calibration will continue. 19. Once the GAS 2 cal
Calibrating an Instrument with Special Sensors • 62EDIT ENTER button until the home screen appears. • To turn the EAGLE 2 off:If the calibration was
63 • TroubleshootingTroubleshootingNOTE: This troubleshooting guide describes SDM-E2 problems only. See the EAGLE 2 Operator’s Manual for problems yo
Troubleshooting • 64• No IR connection between EAGLE 2 and SDM-E2• SDM-E2 is not turned on.• The EAGLE 2 is not correctly inserted into the instrumen
System Requirements • 2CAUTION: The EAGLE 2 detects oxygen deficiency and elevated levels of oxygen, combustible gases, carbon monoxide, and hydrogen
65 • Charging an Instrument in a Docking StationCharging an Instrument in a Docking StationThe SDM-E2 can be used to charge the rechargeable NiMH bat
Charging an Instrument in a Docking Station • 66NOTE: If alkaline batteries are detected, the SDM-E2 must be turned off and on again before attemptin
67 • Charging an Instrument in a Docking StationBatteries Too Drained for Standalone OperationIf the charger is connected to the EAGLE 2 and the batt
Calibration and Bump Test Records • 68Calibration and Bump Test RecordsThe SDM-E2 saves a record of each bump test and calibration performed. It is c
69 • Calibration and Bump Test RecordsSDM-E2 will take a few seconds to determine how much memory is available in the flash drive. • If the flash dri
Calibration and Bump Test Records • 70The files that an SDM-E2 saves to a USB flash drive are structured so that they can be imported into a database
71 • OverviewChapter 5: Single Module Data Viewer ProgramOverviewThe Single Module Data Viewer Program is used to view, organize, and print bump test
Using the Single Module Data Viewer Program • 72NOTE: When you start the Single Module Data Viewer Program for the first time, there will be no data
73 • Using the Single Module Data Viewer Program• View the bump test and calibration data saved in the database.• Delete data.• Print bump test or ca
Using the Single Module Data Viewer Program • 74Organizing the Data When viewing the data, it can be organized in two ways: 1. Base View FormatIn bas
3 • SpecificationsSpecificationsTable 1: SDM-E2 SpecificationsInput Power 12 VDCNOTE: AC Adapter with 100 - 240 VAC, 50/60 Hz, 0.6A input and 12 VDC,
75 • Using the Single Module Data Viewer Program 2. ID View FormatIn ID view format, the data can be organized by one or more of the following items
Using the Single Module Data Viewer Program • 76 2. When an item no longer has a (+) or (-) symbol next to it, single click it and the contents of th
77 • Using the Single Module Data Viewer ProgramIf you organize the data by date, then folders organized by year/month appear in the left side of the
Using the Single Module Data Viewer Program • 78If you are viewing data in ID view format, expand the folders in the left side of the window until th
79 • Using the Single Module Data Viewer Programappear showing the selected printer and confirming if you want to print. Verify that the selected pri
Spare Parts List • 80window will appear prompting you for the new password again. 5. Enter the new password again and click the Confirm New Password
81 • Spare Parts List81-0018RK-01 Calibration cylinder, 50 %LEL isobutane in air, 34 liter steel81-0018RK-03 Calibration cylinder, 50 %LEL isobutane
Spare Parts List • 8281-0142RK-04 5-gas calibration cylinder, CH4/O2/H2S/CO/SO2, 34 liter aluminum81-0143RK-02 Five-gas calibration cylinder, CH4/O2/
83 • Appendix A: Bump Testing and Calibrating EAGLE 2s with IR SensorsAppendix A: Bump Testing and Calibrating EAGLE 2s with IR SensorsThe default ga
Appendix A: Bump Testing and Calibrating EAGLE 2s with IR Sensors • 84An IR sensor that can be bump tested or calibrated with one or more of the stan
About this Manual • 4About this ManualThe SDM-E2 Docking Station Standalone Configuration Operator’s Manual uses the following conventions for notes,
Comentarios a estos manuales