RKI Instruments, Inc.
Data Cal 2000
Operators Manual
Part Number: 71-0097RK
Revision: D
Released: 11/3/11
Vista de pagina 0
1 2 3 4 5 6 ... 87 88

Indice de contenidos

Pagina 1 - Operator’s Manual

RKI Instruments, Inc. Data Cal 2000Operator’s ManualPart Number: 71-0097RKRevision: DReleased: 11/3/11

Pagina 2 - Warranty

5 • DescriptionDocking StationThe docking station manages the calibration, bump test, and information exchange functions of the Data Cal 2000 once th

Pagina 3 - Table of Contents

Description • 6Back PanelThe power switch is a rocker switch and is located in the upper right corner of the back panel. The docking station is on wh

Pagina 4

7 • Descriptionrecessed plug in the lower right corner of the panel is used to connect an AC line cord. It is labeled AC IN.A fuse holder is located

Pagina 5

Description • 8Two DB-15 connectors on the back of each module allow connection to other docking modules or directly to the docking station.Two hose

Pagina 6 - Introduction

9 • Descriptionmodule is the last one in the flow system, it is plugged with a push-on plug that is supplied with the docking station.A two port exha

Pagina 7 - System Requirements

Description • 10DM-2001 Docking ModuleThe GX-2001 fits into the DM-2001 module as shown in Figure 7 below. The module includes a snap-on sample adapt

Pagina 8 - Specifications

11 • DescriptionDM-2003 Docking ModuleThe GX-2003 fits into the module as shown in Figure 8 below. A coiled black tube with a plastic push-on fitting

Pagina 9 - Description

Description • 12DM-2009 Docking ModuleThe GX-2009 fits into the module as shown in Figure 9 below. Communication CablesThe number of communication ca

Pagina 10 - Docking Station

13 • Data Cal 2000 Hardware SetupPower CordA three-pronged AC power cord is provided with the Data Cal 2000. One end connects to a standard AC wall o

Pagina 11 - Description • 6

Data Cal 2000 Hardware Setup • 14Setting the Docking Module Configuration SwitchesThe docking module configuration switches determine module ID and t

Pagina 12 - Cal 2000

WarrantyWarrantyRKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials and workmanship, and performance f

Pagina 13

15 • Data Cal 2000 Hardware SetupTermination SwitchThe termination switch is located above the left DB-15 connector on the back of the docking module

Pagina 14 - 9 • Description

Data Cal 2000 Hardware Setup • 16 3. Place the docking station on the flat surface with enough room to the right of the docking station (when viewed

Pagina 15 - DM-2001 DM-2001

17 • Data Cal 2000 Hardware Setupused in your system, cable installation is done. If more modules are used in you system, continue with step 4. 4. In

Pagina 16 - To GX-2003 Exhaust Fitting

Data Cal 2000 Hardware Setup • 18 3. Repeat step 2 for each successive module in the system until you get to the last module. 4. Install the black fi

Pagina 17 - Communication Cables

19 • Data Cal 2000 Hardware SetupCAUTION: The maximum recommended length for the exhaust tube is 10 feet. Do not use more than 10 feet of tubing for

Pagina 18 - Data Cal 2000 Hardware Setup

Data Cal 2000 Hardware Setup • 20Docking Station Calibration Gas Fitting ConnectionsThe calibration gas fittings on the back panel of the docking sta

Pagina 19

21 • Installing the Data Cal 2000 Docking Station Software 1. If the area around the calibration station is not considered a fresh air area, an area

Pagina 20 - Mechanical Assembly

Installing the Data Cal 2000 Docking Station Software • 22 5. Read the license agreement and click the agreement acceptance selection box, then click

Pagina 21 - Cable Installation

23 • Launching the Data Cal 2000 SoftwareLaunching the Data Cal 2000 Software 1. Click Start on the Windows® Icon Tray, then select Programs/Docking

Pagina 22 - Tubing Installation

Overview of the Data Cal 2000 Software • 24Cal 2000 Software".If you have already performed the software setup, proceed to “Connecting Instrumen

Pagina 23 - GX-2001 and/or GX-2003

Table of ContentsTable of ContentsIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Syste

Pagina 24 - AIR GAS 1 GAS 2

25 • Overview of the Data Cal 2000 Software• Instrument calibration controlled by the Calibration buttonMain Program WindowThe main program window is

Pagina 25

Overview of the Data Cal 2000 Software • 26Docking Module/Instrument Display AreaThe large middle portion of the main window displays docking modules

Pagina 26 - Station Software

27 • Setting Up the Data Cal 2000 Software• Orange indicates that the corresponding cylinder will expire in less than 10 days. When the solenoid is a

Pagina 27

Setting Up the Data Cal 2000 Software • 28Setting Up the Configuration 1. Launch the software as described in “Launching the Data Cal 2000 Software”

Pagina 28

29 • Setting Up the Data Cal 2000 Software 4. Select the communication port that will be used in the Communication section. This is typically COM1 or

Pagina 29 - Data Cal 2000 Functions

Setting Up the Data Cal 2000 Software • 303600 seconds (1 hour). 8. Select the datalogging options in the Data Logger section. • If Down Load Data Lo

Pagina 30 - Main Program Window

31 • Setting Up the Data Cal 2000 Softwareto 90 seconds• Set the bump test tolerance for each channel. This option for each channel allows you to set

Pagina 31

Setting Up the Data Cal 2000 Software • 32Viewing the Pre-Defined Cylinders 1. From the main program window, click the Cylinders button located along

Pagina 32 - Software

33 • Setting Up the Data Cal 2000 SoftwareAdding and Updating CylindersAdd a cylinder if the correct cylinder for your bump testing and calibration n

Pagina 33 - Setting Up the Configuration

Setting Up the Data Cal 2000 Software • 34Adding a Cylinder to the Cylinder List 1. From the Cylinders window, click the New button. The CylinderDeta

Pagina 34

Table of ContentsSetting Up the Data Cal 2000 Software . . . . . . . . . . . . . . . . . . . . . . . . 27Setting Up the Configuration . . . . . .

Pagina 35

35 • Setting Up the Data Cal 2000 Software 7. For each additional gas in the cylinder being entered move to the next Symbol(Unit) field and repeat st

Pagina 36

Setting Up the Data Cal 2000 Software • 36 3. Update the cylinder parameters as desired. 4. Click the OK button. A confirmation window appears. 5. Cl

Pagina 37

37 • Setting Up the Data Cal 2000 Software• The GAS 1 solenoid which controls the Gas 1 fitting is assigned to Cylinder 0001 which is RKI 4-gas mix 8

Pagina 38 - Adding and Updating Cylinders

Connecting Instruments to the Data Cal 2000 • 38Connecting Instruments to theData Cal 2000When an instrument is connected to the Data Cal 2000, the s

Pagina 39

39 • Connecting Instruments to the Data Cal 2000NOTE:As soon as a DM-2003 docking module recognizes that a GX-2003 is installed, a solenoid in the DM

Pagina 40

Connecting Instruments to the Data Cal 2000 • 40• A GX-2003 will begin its normal startup sequence and after a few seconds, the pump will turn off an

Pagina 41 - Setting Up the Solenoids

41 • Connecting Instruments to the Data Cal 2000 8. Once all instruments are connected and all automatic functions are completed, the instruments are

Pagina 42

Bump Testing Instruments • 42Bump Testing InstrumentsThe Data Cal 2000 is capable of performing a bump test both manually and automatically when an i

Pagina 43 - Connecting Instruments to the

43 • Bump Testing Instruments 3. Click the Bump test button. The Data Cal 2000 will initiate a bump test. During the bump test, the Data Cal 2000 wil

Pagina 44 - PC

Bump Testing Instruments • 44If the instrument or instruments failed the bump test, then the following indication is given. 5. If the Force Calibrati

Pagina 45

Table of ContentsPower Off Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77Configuration Window . . . . . . . .

Pagina 46

45 • Calibrating InstrumentsCalibrating InstrumentsThe Data Cal 2000 is capable of performing a calibration both manually and automatically on an ins

Pagina 47 - Bump Testing Instruments

Calibrating Instruments • 46 4. When the calibration is completed, the software will analyze the results and display the Calibration & Bump test

Pagina 48 - 43 • Bump Testing Instruments

47 • Calibrating InstrumentsIf the instrument or instruments failed the calibration, then the following indication is given. 5. To close the result w

Pagina 49 - Bump Testing Instruments • 44

Deleting Instruments From the Main Window • 48Deleting Instruments From the Main WindowAn instrument, along with all its associated data, can be dele

Pagina 50 - Calibrating Instruments

49 • Charging Instruments With The Data Cal 2000 2. The LED on the left front of the docking module will turn off. The LED on the right front of the

Pagina 51 - Calibrating Instruments • 46

Logs Window • 50NOTE: If the GX-2009 is left in the DM-2009 for a long period of time, the dock pump will run for 12 seconds every 2 minutes in order

Pagina 52 - 47 • Calibrating Instruments

51 • Logs Window 2. ID View FormatThe data can be organized by one or more of the following items depending on which selection box or boxes in the lo

Pagina 53 - Charging Instruments With The

Logs Window • 52Logs window. If an item is expanded and you want to close it, click the (-) symbol next to the item or double click it. 2. When a fol

Pagina 54

53 • Logs Windowbefore an alarm event or is not on for 15 minutes after the alarm event, only the time that the unit was on is shown in the alarm tre

Pagina 55 - Logs Window

Logs Window • 54If you organize the data by date, then folders organized by year/month appear in the left side of the window. Expand the folder you w

Pagina 56 - 51 • Logs Window

1 • IntroductionIntroductionThe Data Cal 2000 Docking Station is a comprehensive and reliable system that provides charging, data downloading, config

Pagina 57 - Alarm Trend Data

55 • Logs Window 3. To view summary information about the alarm file, select it in the right side of the window. A summary of the file will appear in

Pagina 58 - 53 • Logs Window

Logs Window • 56 5. You can choose to view the data in a variety of arrangements, such as data on the left and graph on the right, by clicking the ap

Pagina 59 - Logs Window • 54

57 • Logs Windowselected, the data will print as a table. 8. To save the data to a file which can be imported into another program, select either Lis

Pagina 60 - 55 • Logs Window

Logs Window • 58Calibration DataThe GX-2003 saves the calibration information for the last 20 calibrations while the GX-2009 saves the calibration in

Pagina 61 - Logs Window • 56

59 • Logs WindowIf you organize the data by date, then folders organized by year/month appear in the left side of the window. Expand the folder you w

Pagina 62 - 57 • Logs Window

Logs Window • 60 3. Files that record a failed calibration on all channels are highlighted in red. Files that record a calibration where not all chan

Pagina 63 - Calibration Data

61 • Logs WindowEvent DataThe GX-2003 and the GX-2009 save events such as gas alarms, calibration failures, and sensor failures. These events are sav

Pagina 64 - 59 • Logs Window

Logs Window • 62Expand the Event folder. Folders organized by year/month will be listed below the Event folder. Click the folder whose contents you w

Pagina 65 - Logs Window • 60

63 • Logs Windowbe adjusted with the DataCal 2000 software in the Parameter Tab as described in “Setting Up the Configuration” on page 28. At the sch

Pagina 66 - Event Data

Logs Window • 64If you organize the data by date, then folders organized by year/month appear in the left side of the window. Expand the folder you w

Pagina 67 - Interval Trend Data

System Requirements • 2CAUTION: The operator of this instrument is advised that if the equipment is used in a manner not specified in this manual, th

Pagina 68 - 63 • Logs Window

65 • Logs Window 3. To view summary information about an interval trend file, select it in the right side of the window. A summary of the file will a

Pagina 69 - Logs Window • 64

Logs Window • 66 5. You can choose to view the data in a variety of arrangements, such as data on the left and graph on the right, by clicking the ap

Pagina 70 - 65 • Logs Window

67 • Logs WindowGraph is selected, the data will print as a graph and if List is selected, the data will print as a table. 8. To save the data to a f

Pagina 71 - Logs Window • 66

Logs Window • 68Memo FeatureMemos can be created in the Instrument Information window. See “Edit Function, Memo Entry” on page 75. Memos can be used

Pagina 72 - 67 • Logs Window

69 • Logs WindowIf you are viewing data in ID view format, expand the files in the left side of the window until the Memo folder you wish to view is

Pagina 73 - Memo Feature

Logs Window • 70Bump Test FilesThe results of all bump tests performed by the Data Cal 2000 are saved in the Logs window. To view and perform desired

Pagina 74 - 69 • Logs Window

71 • Logs Windowmonth will be listed below the Bump test folder. Click the folder whose contents you want to view and the bump test files in it will

Pagina 75 - Bump Test Files

Instrument/Module Function Menu • 72Instrument/Module Function MenuWhen you are in the main program window, if you right click on an instrument or a

Pagina 76 - 71 • Logs Window

73 • Instrument/Module Function MenuOpen FunctionThis function can be activated by either right clicking an instrument and selecting Open(O) when the

Pagina 77

Instrument/Module Function Menu • 74mode (see below).Edit FunctionThis function can be activated by either right clicking an instrument that is conne

Pagina 78 - Open Function

3 • SpecificationsSpecificationsTable 1: Data Cal 2000 SpecificationsInput Power 115 VAC, 50/60 HzEnvironmental Conditions• For Indoor Use Only• -10°

Pagina 79 - Edit Function

75 • Instrument/Module Function MenuLoad User Defaults buttons in the Sensor Setting tab. To change any of the items in either of the two tabs’ field

Pagina 80 - Edit Function, Memo Entry

Instrument/Module Function Menu • 76document. 3. Click the OK button in the lower right corner of the Instrument Information window. 4. A confirmatio

Pagina 81 - Reset Function

77 • Configuration WindowNOTE: If communication errors for a module or improper operation of a module continue after resetting the module, verify tha

Pagina 82 - Configuration Window

Configuration Window • 78 2. Enter the password and click OK. The factory set password is “ABCDE” and is case sensitive. After clicking OK, the Confi

Pagina 83 - Parameter Tab

79 • Configuration WindowDatabase Tab 1. From the Configuration window, click on the Database tab to view it. 2. To create a new database that is a c

Pagina 84 - Database Tab

Configuration Window • 80click the Remove selection box. When you have made all selections, click the Export button. A Save As dialog box will appear

Pagina 85 - Sensor Tab

81 • Configuration WindowFollower setting from the factory default settings. Do not change the Zero Follower settings. 4. To save any changes made in

Pagina 86 - Password Tab

Optimizing the Database • 82 5. A confirmation window appears. Click the Yes button. You will return to the main program window.Optimizing the Databa

Pagina 87 - Spare Parts List

83 • Spare Parts List81-DM2001 Docking module for GX-200181-DM2003 Docking module for GX-200381-DM2009 Docking module for GX-2009Table 3: Spare Parts

Pagina 88 - Table 3: Spare Parts List

Description • 4DescriptionThe Data Cal 2000 consists of the following components:• Data Cal 2000 Docking Station• DM-2001 Docking Module for GX-2001•