www.rkiinstruments.com
GX-2009
Portable Gas Monitor
Operators Manual
Part Number: 71-0158RK
Revision: H
Released: 6/18/14
Vista de pagina 0
1 2 3 4 5 6 ... 72 73

Indice de contenidos

Pagina 1 - Operator’s Manual

www.rkiinstruments.comGX-2009 Portable Gas MonitorOperator’s ManualPart Number: 71-0158RKRevision: HReleased: 6/18/14

Pagina 2 - GX-2009 Operator’s Manual

5 • Description GX-2009 Operator’s ManualCaseThe GX-2009’s sturdy, high-impact plastic case is suitable for use in many environmental conditions, ind

Pagina 3 - Warranty

GX-2009 Operator’s Manual Description • 6Alligator & Belt ClipsThe GX-2009 is available with two types of clips, the standard alligator clip and

Pagina 4 - Table of Contents

7 • Description GX-2009 Operator’s Manualsensor will respond if exposed to H2S and certain hydrocarbon gasses. The charcoal filter scrubs these gasse

Pagina 5

GX-2009 Operator’s Manual Description • 8Oxygen (O2) SensorThe O2 sensor is a galvanic type sensor. A membrane covers the cell and allows gas to diff

Pagina 6 - Introduction

9 • Description GX-2009 Operator’s ManualPrinted Circuit BoardsThe primary function of the GX-2009’s printed circuit boards is to amplify the signal

Pagina 7 - Specifications

GX-2009 Operator’s Manual Description • 10charge remaining in the batteries. See “Recharging the GX-2009’s NiMH Batteries” on page 55 for instruction

Pagina 8

11 • Description GX-2009 Operator’s ManualCharging stations for more than one instrument include additional chargers that are attached to each other.

Pagina 9 - Description

GX-2009 Operator’s Manual Description • 12DC Powered Charging StationAn optional DC powered single instrument charging station is available with a ve

Pagina 10 - Battery Charging Contacts

13 • Start Up GX-2009 Operator’s ManualStart UpThis section explains how to start up the GX-2009 and get it ready for operation.Start-up Procedure1.

Pagina 11 - Scrubber Filters/Sensor Cover

GX-2009 Operator’s Manual Start Up • 14• If calibration is not due or Cal. Limit Check is set to None, then the following screen appears for a few se

Pagina 12

GX-2009 Operator’s ManualWARNINGRead and understand this instruction manual before operating instrument. Improper use of the gas monitor could result

Pagina 13 - Control Buttons

15 • Start Up GX-2009 Operator’s ManualWARNING:You must perform a bump test in order to continue to normal operation. If you do not perform a bump te

Pagina 14 - Printed Circuit Boards

GX-2009 Operator’s Manual Start Up • 168. If the GX-2009 experiences a sensor failure during start up, a screen indicating which sensor failed displa

Pagina 15 - Charging Station

17 • Start Up GX-2009 Operator’s ManualPerforming a Fresh Air AdjustmentBefore using the GX-2009, set the fresh air reading. Performing this adjustme

Pagina 16 - AC Adapter

GX-2009 Operator’s Manual Operation • 18OperationThis section describes the operation of the GX-2009 in Measuring Mode. It explains how enter Display

Pagina 17

19 • Operation GX-2009 Operator’s Manualand TWA values are cleared when the unit is turned off.To enter Display Mode and view items or reset the peak

Pagina 18 - C--LIMIT

GX-2009 Operator’s Manual Operation • 207. Press and release the POWER MODE button again to return to Measuring Mode.NOTE: If you do not press a butt

Pagina 19 - BUMP NOW

21 • Operation GX-2009 Operator’s Manual• The GX-2009 provides the combustible sensor with some protection against exposure to high levels of combust

Pagina 20 - NEXT BUMP

GX-2009 Operator’s Manual Operation • 22Alarm IndicationsThe GX-2009 will sound an alarm, the LEDs will flash, and the unit will vibrate when one of

Pagina 21

23 • Operation GX-2009 Operator’s ManualAlarm Logic/Resetting AlarmsThe gas alarm logic can be set to either latching or self resetting with the GX-2

Pagina 22 - A i r C A L

GX-2009 Operator’s Manual Operation • 24alarm setpoint (or above for oxygen Warning).If the alarms are self-resetting, then an alarm condition will a

Pagina 23 - 0 0

GX-2009 Operator’s Manual WarrantyWarrantyRKI Instruments, Inc. warrants the GX-2009 sold by us to be free from defects in materials, workmanship, an

Pagina 24 - HOLD AIR

25 • Operation GX-2009 Operator’s ManualResponding to Battery AlarmsWARNING: The GX-2009 is not operational as a gas monitoring device during a dead

Pagina 25 - Combustible Gas Detection

GX-2009 Operator’s Manual Operation • 261. Turn on the GX-2009 as directed in “Start Up” on page 13.2. Attach the sample adapter plate to the sensor

Pagina 26

27 • Data Logging GX-2009 Operator’s ManualNOTE: The peak readings for each channel are saved until a higher peak is recorded, the peak readings are

Pagina 27

GX-2009 Operator’s Manual Calibration Mode • 28Calibration ModeThis section describes the GX-2009 in Calibration Mode. In Calibration Mode, you can m

Pagina 28

29 • Calibration Mode GX-2009 Operator’s ManualUsing Calibration ModeWARNING: The GX-2009 is not in operation as a gas detector while in Calibration

Pagina 29

GX-2009 Operator’s Manual Calibration Mode • 306. When you are done using the menu items in Calibration Mode, use the AIR button to scroll through th

Pagina 30 - Aspirator Adapter

31 • Calibration Mode GX-2009 Operator’s ManualCAL menu item to perform a span adjustment. A fresh air adjustment performed in Calibration Mode is th

Pagina 31 - Sample Adapter

GX-2009 Operator’s Manual Calibration Mode • 32calibration kit. The procedure below describes a span adjustment of all four channels using a calibrat

Pagina 32 - GX-2009 User Setup Program

33 • Calibration Mode GX-2009 Operator’s Manualthe AIR button.• When the desired channel is displayed, press and release the POWER MODE button to sel

Pagina 33 - Calibration Mode

GX-2009 Operator’s Manual Calibration Mode • 343. Use the sample tubing to connect the calibration adapter plate to the regulator. Attach the tubing

Pagina 34 - Using Calibration Mode

Table of Contents GX-2009 Operator’s ManualTable of ContentsIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 35 - 08 - 8

35 • Calibration Mode GX-2009 Operator’s Manual6. Press and release the POWER MODE button. The LCD will display the current gas readings and “AUTO CA

Pagina 36 - AUTO CAL

GX-2009 Operator’s Manual Calibration Mode • 3617.Store the components of the calibration kit in a safe and convenient place.Performing a Span Adjust

Pagina 37 - 50 25

37 • Calibration Mode GX-2009 Operator’s Manualdescribed in “Performing a Fresh Air Adjustment” on page 30.2. At the ONE CAL screen, press and releas

Pagina 38

GX-2009 Operator’s Manual Calibration Mode • 386. Push the adapter plate onto the GX-2009’s sensor face. Make sure the adapter plate is oriented as s

Pagina 39 - Adapter Plate

39 • Calibration Mode GX-2009 Operator’s Manual12.The GX-2009 will make the span adjustment and will return to the channel selection screen.The chann

Pagina 40

GX-2009 Operator’s Manual Calibration Mode • 40Performing a Bump Test in BUMPNOTE: Bump Test Function must be set to On using the GX-2009 User Setup

Pagina 41 - ONE CAL

41 • Calibration Mode GX-2009 Operator’s ManualNOTE: The bump test gas concentrations are the same as the AUTO CAL gas concentrations.If the values a

Pagina 42

GX-2009 Operator’s Manual Calibration Mode • 425. Push the adapter plate onto the GX-2009’s sensor face. Make sure the adapter plate is oriented as s

Pagina 43

43 • Calibration Mode GX-2009 Operator’s Manual9. When Calibration After Bump Test Failed is set to OFF:• The instrument will display which channels

Pagina 44

GX-2009 Operator’s Manual Calibration Mode • 4410.When Calibration After Bump Test Failed is set to ON:If all channels pass the bump test, • The foll

Pagina 45 - BUMP

GX-2009 Operator’s Manual Table of ContentsUser Setup Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 46

45 • Calibration Mode GX-2009 Operator’s Manual• To view the calibration gas reading, press the AIR button again.• To return to the BUMP screen in Ca

Pagina 47

GX-2009 Operator’s Manual Calibration Mode • 461. When the REFRESH screen is displayed, press and release the POWER MODE button. Two screens begin al

Pagina 48 - BUMP

47 • User Setup Mode GX-2009 Operator’s ManualUser Setup ModeThis section describes the GX-2009 in User Setup Mode. User setup mode has the same menu

Pagina 49

GX-2009 Operator’s Manual User Setup Mode • 482. Press and hold the AIR button, then press and hold the POWER MODE button. You will hear a beep after

Pagina 50 - 48 0

49 • User Setup Mode GX-2009 Operator’s Manualadjustment. To switch from increasing to decreasing a value or decreasing to increasing a value, do the

Pagina 51 - TURN OFF

GX-2009 Operator’s Manual User Setup Mode • 50Setting the Alarm PointsEntering the ALARM--P menu item allows you to set the alarm points for each cha

Pagina 52 - User Setup Mode

51 • User Setup Mode GX-2009 Operator’s Manual3. When the desired channel is displayed, press and release the POWER MODE button to select the channel

Pagina 53 - PASSWORD

GX-2009 Operator’s Manual User Setup Mode • 52Setting the PasswordYou can password protect entry into the User Setup Mode with the password feature.

Pagina 54 - Setting the Date and Time

53 • User Setup Mode GX-2009 Operator’s Manual8. Repeat Step 5 through Step 7 until you have set all four of the digits. 9. When you save the last di

Pagina 55 - ALARM--P

GX-2009 Operator’s Manual Maintenance • 54MaintenanceThis section describes troubleshooting procedures for the GX-2009. It also describes how to rech

Pagina 56

1 • Introduction GX-2009 Operator’s ManualIntroductionUsing an advanced detection system consisting of four gas sensors, the GX-2009 personal four-ga

Pagina 57

55 • Maintenance GX-2009 Operator’s ManualRecharging the GX-2009’s NiMH BatteriesWARNING: To prevent ignition of a hazardous atmosphere, batteries mu

Pagina 58

GX-2009 Operator’s Manual Maintenance • 563. If using an AC powered charging station, plug the AC adapter into an electrical outlet.To AC OutletAC Po

Pagina 59 - Maintenance

57 • Maintenance GX-2009 Operator’s ManualIf using a DC powered charging station, plug the 12 VDC vehicle plug adapter into a vehicle’s 12 VDC power

Pagina 60

GX-2009 Operator’s Manual Maintenance • 584. Put the GX-2009 in the charging base. Tabs at the top and bottom of the charging base retain the instrum

Pagina 61 - AC Powered Charging Station

59 • Maintenance GX-2009 Operator’s Manual“memory effect” and the run time of the instrument may be decreased. If you notice this effect with your in

Pagina 62 - DC Powered Charging Station

GX-2009 Operator’s Manual Maintenance • 605. As you pull the top of the rear case away from the front case, you will feel a slight click as the botto

Pagina 63 - SCO%LELO

61 • Maintenance GX-2009 Operator’s Manual8. Replace each battery with a new one. Note the polarity markings on the circuit board and batteries when

Pagina 64 - Pull rear case away

GX-2009 Operator’s Manual Maintenance • 62Replacing the Scrubber FiltersWARNING: Replace the scrubber filters in a non-hazardous environment.1. Verif

Pagina 65 - Figure 21: NiMH Batteries

63 • Maintenance GX-2009 Operator’s Manual7. Reattach the sensor cover retainer in its original position. It should snap into place onto the sensor r

Pagina 66

GX-2009 Operator’s Manual Maintenance • 644. Remove the sensor from its socket.5. Carefully insert the replacement sensor in the socket.POWER MODEAIR

Pagina 67 - H2SCO%LELO2

GX-2009 Operator’s Manual Specifications • 2SpecificationsTable 1: Model GX-2009 SpecificationsTarget Gas Combustibles,Methane (CH4) calibration stan

Pagina 68 - Replacing a Sensor

65 • Maintenance GX-2009 Operator’s ManualCAUTION: When replacing a sensor, verify that the sensor is properly aligned with its socket before inserti

Pagina 69 - RKI GX-2009

GX-2009 Operator’s Manual Parts List • 66Parts ListTable 7 lists replacement parts and accessories for the GX-2009.Table 7: Parts ListPart Number Des

Pagina 70

67 • Parts List GX-2009 Operator’s Manual49-2170RK-04 Charging station with AC adapter for four instruments, 115 VAC49-2170RK-05 Charging station wit

Pagina 71 - Parts List

GX-2009 Operator’s Manual Parts List • 6881-1004RK Regulator, fixed flow, 0.5 LPM, 58/34 liter aluminum, 103 liter steel81-1107RK Calibration/sample

Pagina 72 - Part Number Description

3 • Specifications GX-2009 Operator’s ManualNOTE: The bump test function in Calibration Mode is included in instruments with firmware version 03955 o

Pagina 73

GX-2009 Operator’s Manual Description • 4DescriptionThis section describes the GX-2009’s components and the charging station. These components includ