5 How to Operate 5-7. User mode
RM-5000 - 38 -
5-7. User mode
<<User Mode>>
WARNING
After the adjustment is completed, never fail to press the MODE key to return to the detection mode.
(If the indicator/alarm unit remains in the user mode, it automatically returns to the detection mode
in ten hours.)
Detection Mode
Press the MODE key for three seconds.
User Mode
1-1. ZERO
Perform the zero adjustment.
(In case of OX-5001 and OX-5002,
perform the span adjustment.)
Zero Adjustment => P40
1-2.CONFIRM
Show the setting of the typical
menu.
Setting Display => P41
1-3. PEAK
Displays the maximum
concentration value when a gas
is detected.
1-4. RM VER
Show the program version of the
main unit.
1-5. AMP VER
Show the program version of the
A
MP unit.
1-6. ADDRESS
Show the communication
address of RS-485.
(Only on a
model with an option mounted)
During maintenance
During inhibit
Lamp on
Lamp off
Lamp blinking