Rki-instruments RM-5000 Series Indicator/Alarm Unit Manual de usuario Pagina 55

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 81
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 54
7 Maintenance 7-2. Regular maintenance mode
- 55 - RM-5000
<<Regular Maintenance Mode>>
During maintenance
During inhibit
Lamp on
Lamp off
Lamp blinking
User Mode
In "1-8.M MODE", press the
SET key.
Then press the SET key
again for three seconds.
Regular Maintenance Mode
2-0. GAS TEST
Perform a test with the gas.
Similar to the detection
mode, the reading changes
and the alarm lamp lights up
after the gas is introduced,
but the contact is not
activated.
(Alternate Display)
2-1. ZERO
Perform the zero adjustment.
Zero Adjustment
=> P69
2-2. SPAN
Perform the span
adjustment.
Span Adjustment
=> P70
2-3. LAST CAL
Show the last calibrated date.
(Example: 2011/01/01 12:00)
2-4. HEAT CUR
Display the value of a current
that runs in the heater of the
sensor.
(GP-5001, NC-5001,
NC-5001W, GH-5001 only)
2-5. SETTING1
Specify the environmental
setting 1.
Environmental Setting 1
=> P57
Vista de pagina 54
1 2 ... 50 51 52 53 54 55 56 57 58 59 60 ... 80 81

Comentarios a estos manuales

Sin comentarios