GX-2012 Operator’s Manual Using Maintenance Mode 137
Updating the Leak Check/Bar Hole Mode Setting
With L./B. MODE turned OFF (factory setting), the GX-2012 will
automatically enter Normal Mode when it is turned on.
With L./B. MODE set to bH, you will be prompted to choose between
Normal Mode and Bar Hole Mode when the GX-2012 is turned on.
With L./B. MODE set to LC, you will be prompted to choose between
Normal Mode and Leak Check Mode when the GX-2012 is turned on.
With L./B. MODE set to LC bH, you will be prompted to choose between
Normal Mode, Bar Hole Mode, and Leak Check Mode when the GX-2012 is
turned on.
1. Use the AIRS or (SHIFT)T buttons to display the L./B. MODE menu
item.
2. Press and release the POWER ENTER button. The screen will indicate
the current setting.
3. Use the AIRS or (SHIFT)T button to display the desired setting.
4. Press and release the POWER ENTER button to save the setting and
return to the main menu.
MAINTENANCE
L./B. MODE
OFF
MAINTENANCE
L./B. MODE