Rki-instruments GX-2012 Manual Manual de usuario Pagina 63

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 185
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 62
GX-2012 Operator’s Manual Display Mode • 55
Peak Bar Select Screen
The Peak Bar Select Screen allows you to turn the peak bar on or off. If the
function is turned on, the peak bar appears along the left side of the screen
during Normal Operation and on the Peak Screen in Display Mode. It shows
the peak readings for each channel in bar graph format. The bottom line of
this screen alternates between the current setting, NO /DISP, and YES/ENT.
To adjust the peak bar selection:
1. Use the DISPLAY (ADJ) button to scroll to the Peak Bar Select Screen.
The screen indicates the current setting.
2. Press and release the POWER ENTER button.
3. Use the AIRS and (SHIFT)T buttons to make your selection and press
and release the POWER ENTER button.
bAr
SEL
BAR OFF
bAr
SEL
BAR ON
On
PEAK BAR