56 Data Logging GX-2012 Operator’s Manual
Data Logging
NOTE: The GX-2012 logs data in both Normal Mode and Bar Hole Mode.
The GX-2012 features the ability to log data to its internal memory and
download it to a computer via the infrared communications port located just
below the RESET SILENCE button. It logs gas readings during normal
operation, alarm data, and calibration data.
NOTE: The data logging function is separate from the snap logging
function. For a description of the snap logging function, see “Snap
Log Mode” on page 33.
To utilize the GX-2012’s downloading capability, you will need the GX-2012
downloading software and a computer with an infrared port or a USB port
that runs one of the following operating systems: Windows XP, Windows
Vista, or Windows 7. If your computer has an infrared port, then no
additional accessories are needed to download data from the GX-2012. If
your computer does not have an infrared port but does have a USB port, a
USB/IrDA adapter cable can be used to download data from the GX-2012
using a USB port. The downloading software is available from RKI
Instruments, Inc. The adapter cable is also available from RKI or may be
purchased at a computer or electronics supply store.
For a complete description of the downloading software and procedures for
downloading data to a computer, see the GX-2012 Data Logger Management
Program Operators Manual.
SDM-2012
The GX-2012 can be used with the SDM-2012 if desired. The
SDM-2012 can charge the instrument if the lithium ion battery pack is being
used. It can also bump test, calibrate, and retrieve saved data. The SDM-2012
can be operated as a single, stand-alone unit or can be used with up to 9 other
SDM-2012s and operated using a computer. For more information about the
operation of the SDM-2012, see the SDM-2012 operators manual.