72 Calibration Mode GX-2012 Operator’s Manual
To return to the BUMP screen in Calibration Mode, press POWER
ENTER at any time. If any channel failed the calibration, the buzzer
will sound a double pulsing tone until you press and release the
POWER ENTER button.
Disconnect the tubing from the probe.
12. Unscrew the demand flow regulator from the calibration cylinder.
13. With the BUMP menu option displayed, press the (SHIFT)T button until
the NORMAL menu option is displayed.
14. Press and release the POWER ENTER button to return to Measuring
Mode.