Rki-instruments SDM-2012 PC Controlled Configuration Manual de usuario Pagina 194

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 218
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 193
189 • Bump Testing GX-2012s and Gas Tracers
17. After calibration gas has been applied to the ppm sensor, fresh air will be
applied to the standard sensors (catalytic LEL, O
2
, H
2
S (GX-2012 only), and
CO) and the %volume sensor (if installed) for 45 seconds and the PC
Controller program will perform a zero adjustment.
Figure 149: Air Flowing to Standard Sensors
The PC Controller Program will analyze the zero adjustment results. If any of
the standard sensors or the %volume sensor (if installed) failed the zero
adjustment, the bump test will be aborted for that instrument.
Vista de pagina 193
1 2 ... 189 190 191 192 193 194 195 196 197 198 199 ... 217 218

Comentarios a estos manuales

Sin comentarios