59 • Bump Testing a GX-2012
13. If you do not have a %volume sensor installed in your GX-2012(s), continue
with step 14.
If you have a %volume sensor installed, GAS 2 calibration gas will be applied
to the instrument(s) for the duration specified in the Gas Exposure Time
parameter in the Configuration Window.
Figure 42: GAS 2 Calibration Gas Flowing
The PC Controller Program will analyze the results. If the bump test on the
%volume sensor fails and Force Calibration is selected, a calibration will
automatically begin and GAS 2 calibration gas will continue to flow.