211 • Calibrating GX-2012s and Gas Tracers
23. After the fresh air purge is complete,
If the calibration passes, the following window will open indicating that. The
CAL S LED on the SDM-2012 control panel will be solid green. Click OK to
return to the main program window. The CAL S LED will turn off.
Figure 169: Calibration Passed
The instrument will automatically be shut off 20 seconds after a successful
calibration.