198 ESM-01 Calibration EAGLE 2 Operator’s Manual
33. If the ESM-01 channel passed calibration, the following screen sequence
occurs.
If the ESM-01 channel cannot be adjusted to the proper value, a screen
displays that indicates a calibration failure and indicates that the ESM-01
sensor failed to calibrate.
The buzzer and alarm LEDs activate in a double pulsing pattern. Press and
release the POWER ENTER RESET button to reset the alarm and continue to
the Calibration Mode Screen. Attempt to calibrate the ESM-01 channel
again. If the failure continues, investigate the cause. See “Troubleshooting”
on page 70
34. Disconnect the tubing from the probe.
35. Unscrew the demand flow regulator from the calibration cylinder.
36. Use the RANGE SHIFT button to place the cursor next to the
NORMAL OPERATION menu option, then press and release the
POWER ENTER RESET button to return to Measuring Mode.
AUTO CALIBRATION
DONE
REMOVE GAS
AUTO CALIBRATION
END
CALIBRATION MODE
> AUTO CALIBRATION
SINGLE CALIBRATION
PERFORM AIR ADJUST
NORMAL OPERATION
FAILED SENSOR
ENTER TO CONTINUE
<SO2>