EAGLE 2 Operator’s Manual TC Calibration 209
7. Press and release the POWER ENTER RESET button. The following
screen appears.
8. Press and release the AIR YES button to continue.
If you do not want to continue, press the DISPLAY ADJUST NO button
and the unit will return to the Calibration Mode Screen.
9. The EAGLE 2 will indicate that it is adjusting the zero reading for a few
seconds, then indicate that the operation is complete before returning to
the Calibration Mode Screen.
10. Install the demand flow regulator onto the multi-gas calibration cylinder.
11. Connect the sample tubing to the demand flow regulator.
12. Install the probe on the EAGLE 2 inlet fitting. Make sure the probe is
PERFORM
AIR ADJUST?
ADJUSTING ZERO
RELEASE AIR BUTTON
ADJUSTING ZERO
COMPLETE
CALIBRATION MODE
AUTO CALIBRATION
SINGLE CALIBRATION
> PERFORM AIR ADJUST
NORMAL OPERATION