280 Start Up, Leak Check Mode EAGLE 2 Operator’s Manual
4. The LCD will show the following screen for about ten seconds.
5. The Battery Voltage Screen appears for a few seconds.
6. The Mode Select Screen displays.
7. The cursor will flash in front of NORMAL MODE. Use the
RANGE SHIFT button to move the cursor next to LEAK CHECK
MODE.
8. With Leak Check Mode selected, press and release the POWER
ENTER RESET button to begin the Leak Check Mode startup
sequence.
NOTE: If no button is pressed for 20 seconds, the unit will proceed
into whichever mode has the cursor in front of it.
9. The Active Gases Screen appears for a few seconds indicating which
channels are active and their target gas.
GAS MONITOR
EAGLE 2
BATTERY MIN:
4.3 VOLTS
BATTERY NOW:
5.2 VOLTS
> NORMAL MODE
BAR HOLE MODE
LEAK CHECK MODE
ACTIVE GASES
CH4 OXY
H2S CO
CH4