208 TC Calibration EAGLE 2 Operator’s Manual
appropriate for the TC detection range.
If the EAGLE 2 is intended for use in a landfill, RKI Instruments, Inc.
recommends using the carbon dioxide/methane mix calibration cylinder
because it is representative of gases present in a landfill.
NOTE: If your instrument has more than one TC sensor, you will need a
calibration cylinder for each sensor.
2. Find a fresh-air environment. This is an environment free of toxic or
combustible gases and of normal oxygen content (20.9%).
3. While in Measuring Mode, press and hold the RANGE SHIFT button,
then press the DISPLAY ADJUST NO button and release both buttons.
4. If the unit prompts you for the password, enter it by using the AIR
YES and RANGE SHIFT buttons to select each password number and
then pressing and releasing POWER ENTER RESET to enter the number
and move on to the next one.
5. The Calibration Mode Screen displays with the cursor next to AUTO
CALIBRATION.
6. Move the cursor to the PERFORM AIR ADJUST menu item by using
the RANGE SHIFT button.
CALIBRATION MODE
> AUTO CALIBRATION
SINGLE CALIBRATION
PERFORM AIR ADJUST
NORMAL OPERATION
CALIBRATION MODE
AUTO CALIBRATION
SINGLE CALIBRATION
> PERFORM AIR ADJUST
NORMAL OPERATION