Rki-instruments EAGLE 2 Manual Manual de usuario Pagina 307

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 309
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 306
EAGLE 2 Operator’s Manual Start Up 297
WARNING: The EAGLE 2 is not a gas monitoring device until you select
a mode and enter it.
5. If you select Normal Mode, the warm up sequence will continue as
described in “Start Up” on page 22 unless Leak Check or Bar Hole Mode
is active. If Leak Check or Bar Hole Mode is active, the Leak Check/Bar
Hole Mode Select Screen will appear immediately after the Normal
Mode selection.
Use the AIR YES or RANGE
SHIFT button to display the mode you
wish to enter.
6. If you select Inert Mode, the warm up sequence will continue as
described in “Start Up” on page 22. The oxygen alarms in the alarm point
screens will be the alarm points for Inert Mode and not for Normal Mode.
7. When the warm up is complete and the instrument is operating in Inert
Mode, the screen will appear as shown below if the instrument is
sampling a fresh air environment.
If the instrument is sampling a fresh air environment, the oxygen channel
will be in alarm.
> NORMAL MODE
BAR HOLE MODE
LEAK CHECK MODE
I CH4 0%LEL
N OXY 20.9vol%ALRM2
E H2S 0.0ppm
R CO 0ppm
T
Vista de pagina 306
1 2 ... 302 303 304 305 306 307 308 309

Comentarios a estos manuales

Sin comentarios