Rki-instruments SDM-E2 Manual de usuario Pagina 100

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 224
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 99
95 • Calibrating an Instrument with Special Sensors
12. After the GAS 1 calibration gas is applied, calibration gas for the first special
sensor will then begin flowing. The GAS 2 indicator will come on. Be sure that
the appropriate cylinder is attached to the GAS 2 inlet. The GAS 2 calibration
gas will be applied to the EAGLE 2 for an automatically defined amount of time
based on the type of special sensor.
Figure 72: GAS 2 Flowing
The PC Controller Program will analyze the results.
Vista de pagina 99
1 2 ... 95 96 97 98 99 100 101 102 103 104 105 ... 223 224

Comentarios a estos manuales

Sin comentarios