Rki-instruments SDM-E2 Manual de usuario Pagina 73

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 224
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 72
Calibrating a Standard 4-Gas Instrument • 68
13. After the fresh air purge is complete,
If the calibration passes, the following window will open indicating that. The
CAL S LED on the SDM-E2 control panel will be solid green. Click OK to
return to the main program window. The CAL S LED will turn off.
Figure 48: Calibration Passed
The EAGLE 2 will automatically be shut off 20 seconds after a successful
calibration.
Vista de pagina 72
1 2 ... 68 69 70 71 72 73 74 75 76 77 78 ... 223 224

Comentarios a estos manuales

Sin comentarios