Instrument Function Menu 160
There are limitations for the full scale and increment values that depend on the factor
for both the high range and the low range. Table 4 and Table 5 below list these
limitations.
Table 4: High Range PID
Factor
Full
Scale
(ppm)
Increment
0.25-0.36 1000 1
0.37-0.49 1500 1
0.50-0.61 2000 1
0.62-0.74 2500 1
0.75-0.99 3000 2
1.00-1.24 4000 2
1.25-2.49 5000 2
2.50-3.74 10000 10
3.75-4.99 15000 10
5.00-7.49 20000 10
7.50-9.99 30000 20
10.00-12.49 40000 20
12.50 50000 2500
Table 5: Low Range PID
Factor
Full
Scale
(ppm)
Increment
0.20-0.29 10.00 0.01
0.30-0.39 15.00 0.01
0.40-0.49 20.00 0.01
0.50-0.59 25.00 0.01
0.60-0.79 30.00 0.02
0.80-0.99 40.00 0.02
1.00-1.99 50.00 0.02