89 • Calibrating an Instrument with Special Sensors
7. You will be asked to confirm which EAGLE 2 channels will use GAS 1 for the
calibration. If you have a 5 gas cylinder that includes the target gas of the
special sensor in the appropriate concentration, then be sure that all 5
channels (or fewer if less than 4 standard channels are active) are selected for
GAS 1. If you are using a standard 4-gas cylinder and a separate cylinder for
the special sensor(s), leave the special sensor(s) deselected in this screen.
Figure 66: Gas Cylinder Confirmation Screen
When all of the desired channels are selected, click OK. If you don’t want to
proceed, click Cancel to abort the calibration.
NOTE: If your instrument(s) include one or more infrared sensors, see
“Appendix B: Bump Testing and Calibrating EAGLE 2s with IR
Sensors” on page 218 for a description of the GAS 1/GAS 2
assignments for this type of sensor.
Separate GAS 2 cylinder 5-gas cylinder