Bump Testing Multiple EAGLE 2s with Different Sensor Configurations 188
NOTE: Do not select every gas in the gas cylinder confirmation screen unless
you have calibration cylinders individually connected to your docking
stations. This will result in an inaccurate bump test.
Figure 131: Gas Cylinder Confirmation Screen, Single Channel Special Sensor
When all of the desired channels are selected, click OK. If you don’t want to
proceed, click Cancel to abort the bump test.
NOTE: If your instrument(s) include one or more infrared sensors, see
“Appendix B: Bump Testing and Calibrating EAGLE 2s with IR
Sensors” on page 218 for a description of the GAS 1/GAS 2
assignments for this type of sensor.
No Manifold Assembled Manifold Assembled