Rki-instruments SDM-E2 Manual de usuario Pagina 198

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 224
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 197
193 • Bump Testing Multiple EAGLE 2s with Different Sensor Configura-
12. After air is applied, and if all channels passed the zero adjustment, calibration
gas from the GAS 1 cylinder will be applied. The GAS 1 indicator on the
program screen will come on and calibration gas will be applied to the
instruments for the duration specified in the Gas Exposure Time parameter in
the Configuration Window.
Figure 136: GAS 1 Flowing
The PC Controller Program will analyze the results. If the bump test on the
standard 4 sensors fails and Force Calibration is selected, a calibration will
automatically begin on the standard 4 sensors and GAS 1 calibration gas will
continue to flow.
Vista de pagina 197
1 2 ... 193 194 195 196 197 198 199 200 201 202 203 ... 223 224

Comentarios a estos manuales

Sin comentarios

Alfresco Refrigerator ARFG-42 podręczniki

Instrukcje obsługi i podręczniki użytkownika dla Pomiary, testowanie i kontrola Alfresco Refrigerator ARFG-42.
Dostarczamy 1 pdf podręczników Alfresco Refrigerator ARFG-42 do pobrania za darmo według typów dokumentów: Instrukcja Użytkownika






Więcej produktów i instrukcji dla Pomiary, testowanie i kontrola Alfresco

Modele Rodzaj dokumentu
Brak