Rki-instruments SDM-E2 Manual de usuario Pagina 196

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 224
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 195
191 • Bump Testing Multiple EAGLE 2s with Different Sensor Configura-
If you are bump testing single channel special sensor instruments and have a
manifold assembled, the gas order confirmation screen will appear as shown
below. One sensor will be tested using GAS 1 (in this case, NH
3
). The
remaining sensors will be tested using GAS 2. The GAS 2 shown in the
column directly to the right of the GAS 1 column (in this case, IBL) will always
be bump tested before the GAS 2 shown in the following columns. The GAS 2
selection boxes are deselected by default. You must select the GAS 2 boxes if
you wish to bump test the special sensors. If you do not wish to bump test the
special sensors, leave the GAS 2 boxes deselected. Be sure that the
appropriate GAS 2 cylinder is connected to the GAS 2 fitting. The PC
Controller program will prompt you to change the GAS 2 cylinder during the
bump test.
Figure 134: Gas Order Confirmation Screen, Single Channel Units, Manifold
Vista de pagina 195
1 2 ... 191 192 193 194 195 196 197 198 199 200 201 ... 223 224

Comentarios a estos manuales

Sin comentarios