53 • Bump Testing Standard 4-Gas Instrument(s)
9. You will then be asked to confirm the gas cylinder concentration and the
instruments that the cylinder will be used for. The serial number and gas
combination of all selected instruments will appear sequentially beneath the
gas cylinder concentration.
Figure 34: Gas Order Confirmation Screen
10. Click OK. You will be asked if you want to bump test the instrument(s). Click
OK to proceed. If you don’t want to proceed with the bump test, click Cancel to
return to the main program window.