79 • Bump Testing an Instrument with Special Sensors
13. After GAS 1 calibration gas is applied, calibration gas for the first special
sensor will then begin flowing. The GAS 2 indicator will come on. Be sure that
the appropriate cylinder is attached to the GAS 2 inlet. The GAS 2 calibration
gas will be applied to the instrument(s) for the duration specified in the Gas
Exposure Time parameter in the Configuration Window.
Figure 58: GAS 2 Flowing
The PC Controller Program will analyze the results. If the bump test on the first
special sensor fails and Force Calibration is selected, a calibration will
automatically begin and GAS 2 calibration gas will continue to flow.
Vista de pagina 83