Rki-instruments SDM-E2 Manual de usuario Pagina 57

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 224
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 56
Bump Testing Standard 4-Gas Instrument(s) • 52
8. You will be asked to confirm which EAGLE 2 channels will use GAS 1 for the
bump test. When using a 3-gas (no H
2
S channel) or a 4-gas (H
2
S channel
included) cylinder, to bump test a standard instrument, select all channels. If
your instrument has fewer than 4 channels, your screen will only display the
active channels.
Figure 33: Gas Cylinder Confirmation Screen
When all of the desired channels are selected, click OK. If you don’t want to
proceed, click Cancel to abort the bump test.
Vista de pagina 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 ... 223 224

Comentarios a estos manuales

Sin comentarios