Calibrating Multiple EAGLE 2s with Different Sensor Configurations 210
11. After air is applied, and if all channels passed the zero adjustment, calibration
gas from the GAS 1 cylinder will be applied. The GAS 1 indicator on the
program screen will come on. GAS 1 calibration gas will be applied to the
instrument for 90 seconds.
Figure 151: GAS 1 Flowing
The PC Controller Program will analyze the results.