Rki-instruments SDM-E2 Manual de usuario Pagina 70

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 224
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 69
65 • Calibrating a Standard 4-Gas Instrument
10. The pump will start and the program will indicate that air is flowing. The CAL S
LED will begin to flash amber. Air will be applied to the instrument(s) for 45
seconds, then the PC Controller Program will perform a zero adjustment on all
the channels.
Figure 45: Air Flowing
The PC Controller Program will analyze the zero adjustment results. If any
sensors failed the zero adjustment, the calibration will be aborted.
Vista de pagina 69
1 2 ... 65 66 67 68 69 70 71 72 73 74 75 ... 223 224

Comentarios a estos manuales

Sin comentarios