211 • Calibrating Multiple EAGLE 2s with Different Sensor Configurations
12. After the GAS 1 calibration gas is applied, calibration gas for the first special
sensor will then begin flowing. The GAS 2 indicator will come on. Be sure that
the appropriate cylinder is attached to the GAS 2 inlet. The GAS 2 calibration
gas will be applied to the EAGLE 2 for an automatically defined amount of time
based on the type of special sensor.
Figure 152: GAS 2 Flowing
The PC Controller Program will analyze the results.